Signal intensity scores are a proxy for OCT scan quality and are used commonly in the clinical setting to determine image reliability. Adequate structural illumination, the basis of the signal intensity score, is important for accurate and repeatable OCT measurements, as shown by several studies using the previous generation TD-OCT system.
14–17 It has been reported that higher signal strength results in thicker measurement values
16,17 and this phenomenon also has been shown for RNFL thickness using FD-OCT systems.
18 Figure 1 shows example FD-OCT scans with SSI values above and below the manufacturer's recommendations, and we performed this study to determine the effect of SSI on measurements and their within-visit repeatability. To our knowledge, this study is the first to analyze the effect of SSI on GCC and RNFL measurements using FD-OCT data collected in a prospective longitudinal study.
There are several reasons SSI values may impact measurement values. Optical coherence tomography software algorithms rely on reflectance properties specific to each retinal layer to delineate the inner and outer boundaries of the structure being measured. Software engineers design scan protocols that segment layers with sufficiently different reflectance from their bordering layers, such that the measured layer(s) can be identified by an automated algorithm. When an acquired image has a low overall illumination, which can be caused by media opacities, floaters, cropping, blink artifacts, eye movement, or operator error, the clarity of the image decreases, resulting in less accurate segmentation and increased measurement variability. The directional reflectance of RNFL is another important factor that affects SSI and measured thickness. Along the plane parallel to its length, RNFL has mirror-like directionality,
21 so that the reflected OCT signal is much stronger near perpendicular incidence, and weakens with more oblique incidence angles. Therefore, oblique incidence would reduce SSI through reduced reflectance. It also could artifactually reduce measured RNFL thickness by decreasing the contrast between the usually brighter RNFL and subjacent GCL.
The arrow of causality also could run the other way – a thinner RNFL could lower the SSI of an OCT image because the RNFL usually is one of the more brightly reflecting layer in the peripapillary OCT scan.
Our study found that SSI significantly affects GCC and RNFL measurement repeatability, particularly in scans with SSI values that are on the extreme low end of the spectrum (i.e., SSI < 35). The repeatability of GCC measurements deteriorated more when the SSI scores are very low. It is intuitive that higher quality scans will result in improved measurement repeatability; however, it is important to determine the minimum SSI that generates an acceptable level of repeatability, since, in real world practice, the operator often must accept less than ideal scans. We determined the optimal minimum SSI should be 44 for GCC scans and 37 for RNFL scans. These values result in measurements that are sufficiently repeatable, should be easily obtainable by OCT operators, and further increases in SSI produces only minor improvements in measurement repeatability. In the current study, the best repeatability among RNFL parameters came from scans in SSI bins between 50 and 70; however, only accepting images with these SSI values would result in the exclusion of >30% to 90% of saved scans, and, thus, is an unrealistic requirement for use in clinical practice. Similarly, the best repeatability for GCC scans came from SSI bins between 55 and 70; however, this also represents too strict an SSI range for use in clinical practice.
We also found cropping was a significant source of variability. When the retina is out of view or scan is not centered at designed location, every effort should be made to retake the scan to eliminate these artifacts.
To determine whether SSI affects RNFL thickness, we performed within-visit correlation between variation in SSI and RNFL thickness. This correlation does not measure the effect of RNFL thickness on SSI, since the true RNFL thickness should be constant for the same eye within a visit. Therefore, any correlation would be due to the effect of SSI on RNFL thickness measurement, not the other way around. Using this method, we found a significant correlation between SSI value RNFL measurement thickness among normal and GSPPG eyes (
Table 4). For instance there was a statistically significant (
P < 0.001) 0.056 μm increase in RNFL thickness for each point increase in SSI (SSI scale ranges from 0–100) among normal eyes. This value is less than those found in studies using a TD-OCT device, although it is difficult to make exact comparisons due to different signal intensity scales used by TD-OCT versus RTVue FD-OCT.
13,15,22 Samarawickrama et al.
16 found a small but significant difference in inner macular thickness between scans with moderate and good signal strength scores. These findings suggest that in FD-OCT and TD-OCT devices, weaker SSI is associated with thinner RNFL thickness. The weaker association in later FD-OCT devices could be due to the effort of programmers to reduce this artifactual bias in later generations of segmentation software.
In contrast to RNFL, we did not find a statistically significant correlation between SSI and GCC thickness (
Table 4). This is unusual because GCC and RNFL measure ganglion cells,
23 though in different regions. It would appear that the effect of varying SSI values on measurement thickness does not apply equally to both regions. The reflectance measured by an optical instrument depends not only on the proportion of incident light that is reflected, but also on whether the direction of the reflected light reached the viewing aperture. The directional reflectance of RNFL means that the OCT beam position within the pupil aperture could affect the brightness of RNFL signal on the OCT image. A dim RNFL would reduce the SSI score and reduce the contrast between the RNFL and the GCL, leading to artifactually thinner RNFL measurement. Knighton and Qian
21 have reported the reflectance of the nasal RNFL is particularly sensitive to aperture location and this may account for an increased dependence on adequate signal intensity when scanning a peripapillary pattern. If we hypothesize that the correlation between RNFL thickness measurement and SSI is primarily mediated through variation in OCT beam incidence angle, this could explain why GCC is not affected by SSI as much. The primary segmentation for GCC thickness measurement is the boundary between the IPL and the inner nuclear layer. These layers do not have strong directional reflectance like RNFL, and the signal contrast between them would not be strongly influenced by beam incidence angle.
Another discrepancy was that the within-visit correlation between variations in SSI and RNFL was not found in the PG group, unlike the strong correlation among normal and GSPPG eyes (
Table 4). The reason may be that among PG eyes, the larger loss of nerve fibers or decrease in microtubule content makes reflectance less directional. So change occurs in SSI mainly due to scatter defocus or beam blocking, which does not alter the contrast between RNFL and GCL; therefore, segmentation is not affected. The lack of SSI-RNFL correlation in PG group suggests that direct compensation of SSI for glaucoma progression analysis is not recommended. Although compensating for SSI might reduce variation in RNFL thickness in the normal group, it also would make RNFL appear thicker in the PG group; hence, the contrast between PG and N groups would not be improved by SSI-based compensation. Furthermore, monitoring of RNFL thinning over time as a way of measuring glaucoma progression is useful primarily in the PG group. And SSI-based compensation is not necessary in the PG group.
Several confounding factors, such as age, race, and other variables, are known to affect RNFL and GCC thickness.
17,22,24–28 Older age also had been linked to lower SSI.
26 In the current study, we did not find a significant correlation between age and SSI.
This study has several potential limitations. First of all, the SSI cut-points were determined solely based on repeatability of the measurement under different SSI levels, and the choice of the cut-point is arbitrary, hence there may exist a more optimized cut-point when other factors are considered; also the test comparing repeatability requires same variance across eyes, which may not be met in reality. Secondly, the study used data across multiple tests over multiple visits on both eyes of the participants to compensate for the rather small sample size; correlation and between-visit variation might have a negative effect on the analysis. In an ideal situation, it is preferable to have a large number of same-day repeat sessions on a single eye to evaluate the SSI effect. Additionally, factors other than signal intensity, such as cataract, vitreous floaters, or disc margin delineation algorithm failure,
29 may impact measurement values and lead to increased variability independent of SSI. It also is unknown to what extent measurement variability caused by extreme high and low SSI values affects OCT diagnostic classifications (i.e., green/yellow/red color code), which may complicate interpretation. Finally, although higher SSI values yield improved repeatability, it remains to be determined whether an SSI correction model would improve our ability to track glaucoma progression. Having said that, any advancement that increases our ability to detect true structural changes in GCC and RNFL thickness over time is welcome, and future studies are warranted to determine the role of a SSI thickness-response model in the clinical setting.
In conclusion, higher SSI results in improved repeatability. For the RTVue FD-OCT instrument, we recommend SSI values ≥ 44 for GCC measurements and ≥ 37 for RNFL measurements.