Equipment used includes the following: Shimadzu GC2010/QP2010Plus; column 30 m × 0.25 mm: Phenomenex, Zebron 5%-Phenyl-Aryl-95%-Dimethylpolysiloxane-Phase (1.0 μm).
The qualitative identification of the impurities was carried out through a library spectrum comparison (NIST 21/107).
In addition, impurity groups were identified by extracting the mass traces of selected key fragments of individual impurity types (key fragment CF3 with mass 69 for fluorinated impurities, key fragment COOCH3 with mass 59 for perfluoroalkanoic acid methylesters (methyl-perfluoroalkanoate), key fragment CF2CFO with mass 97 for oxygen-containing fluorinated impurities, key fragment C3F3 with mass 93 for fluorinated alkenes, key fragment CFCOH with mass 61 for fluorinated alcohols and key fragment C6H5CH2 with mass 91 for aromatic compounds (leachables).
Our semiquantitative results after testing impurities relied on evaluating the intensities detected of the related chromatographic peaks and expressed as relative intensities in arbitrary units or in case of the 1H-perfluoroalkane analogues, using the response factor and external standard of 1H-PFO; results are given in ppm.
The quantitative determination was carried out using the method of external standards (1,2 dimethylbenzene, 1,3 dimethylbenzene, 1,4 dimethylbenzene, ethylbenzene, Perfluoro-2-n-butyltetrahydrofurane, methyl-perfluorooctanoate, methyl-perfluorononanoate, PFOA, 1H,1H,7H dodecafluoro-1-heptanol and 1H-PFO).