Advances in microfabrication have allowed the construction of increasingly sophisticated devices well suited to the small dimensions of the eye. Using the technology described above, a high-resolution pressure sensor was integrated with a 30- and 33-gauge needle to accurately and reliably measure IOP in the anterior and vitreous chambers. Notably, the device provides a direct measure of IOP that is not affected by corneal properties. The device accurately measured IOP in the anterior chamber over a clinically significant range of 10 to 100 mm Hg (
Figure 4). In contrast, the Tonopen underestimated the IOP, particularly at higher pressures. This finding is consistent with prior studies showing the Tonopen underestimates IOP in rabbits.
16 IOP measurements in rabbits can be corrected to account for thinner corneas leading to the underestimation of their IOP.
17 Similar correction factors exist for humans, but their use may not lead to increased accuracy in IOP estimation due to many other factors that may induce artifacts.
18 More complex models that attempt to address additional factors such as the modulus of elasticity are still prone to error.
19,20 A history of refractive surgery may lead to further inaccuracies in the measurement of IOP due to thinning of the cornea, changes in the corneal curvature, and alterations in the corneal biomechanical properties.
21–23 Corneal scars may influence IOP in even more unpredictable ways due to their varying sizes, depths, and effects on the cornea's biomechanical properties.
24 All of these potential sources of error are frequently encountered in the clinical setting, yet there are limited means to address them. Our device allows for an accurate measurement of IOP in any of these cases. The patient may not need this measurement repeated at every visit if the results are reassuring or can be correlated to GAT or another noninvasive measurement technique. However, the opportunity for direct IOP measurement would be a useful addition to a clinician's armamentarium.
The device also accurately measured IOP in the vitreous chamber after vitrectomy (
Figure 6). We were unable to measure IOP in the vitreous chamber without vitrectomy because vitreous rapidly clogged the measurement needle, voiding the sensor reading. A similar result was found in prior cannulation studies.
25 However, despite this limitation, direct measurement of IOP in the vitreous chamber after vitrectomy is clinically useful. As many as 60% of Kpro patients develop glaucoma, but the disease is difficult to manage due to the inability to accurately measure IOP.
26 Management of chronic vision-threatening complications like glaucoma in Kpro patients is becoming increasingly important as early complications such as endophthalmitis or device extrusion are becoming less common.
27,28 Many Kpro patients receive vitrectomies at the time of Kpro implantation. These patients may benefit enormously from the accurate measurement of IOP in the vitreous chamber.
Telemetric IOP monitors have been implanted into a small cohort of KPro patients and offers an alternative method for direct measurement of IOP in these patients.
29 However, 3 of 12 devices were explanted over the course of a year, and there were concerns for potential adverse events associated with the devices. Our device may offer a safe alternative in Kpro patients. Interestingly, data from the implantable IOP monitors were compared to anterior chamber manometry.
30 This suggests that it may be possible to measure IOP using our device in KPro patients even without vitrectomy.
However, serial anterior segment imaging has demonstrated progressive angle closure and shallowing of the anterior chamber in KPro patients, so anterior chamber measurements may still not be viable over the long term.
31 Implantable devices also face issues of measurement drift over the lifetime of the device.
32,33 Implantable devices can be re-calibrated to correct for drift by performing GAT in healthy eyes, but this is not possible in KPro patients. Our device may be useful for recalibration of implantable devices as their safety profiles become more acceptable.
The use of the term “gold standard” to describe a diagnostic technique or therapeutic intervention has been criticized as inaccurate or misleading due to the rapidly evolving state of medical care.
34,35 Nonetheless, GAT has long been referred to as the gold standard for IOP measurement.
1 However, accurate measurement of IOP by GAT is hampered by the corneal and biomechanical artifacts discussed above. Anterior chamber cannulation manometry in animal models allows for accurate IOP measurement but was previously hampered by the invasiveness of the technique.
36,37 Now, microfabrication techniques allow clinicians to directly measure IOP through the use of implantable devices or minimally invasive procedures. Thus a true IOP is measured rather than the surrogate IOP measured by noninvasive techniques. The small size of the handpiece and needle make this a feasible clinical measurement, with safety offered by the ability to dispose of the needle and sensor after each use. We propose that these new methods could provide a true gold standard for IOP measurement in appropriately selected patients.
This study had several limitations. First, the study was performed entirely in ex vivo models, so the potential long-term complication rates of direct measurement of IOP in the anterior and vitreous chambers are unknown. However, the safety profiles of anterior chamber paracentesis and intravitreal injections offer promise for a similarly safe procedure that could be performed in an office setting. Second, we performed vitreous chamber measurements in only 2 eyes. The difficulty of fully closing sclerotomies after vitrectomy led to unstable eyes and variable IOP measurements at higher pressures. Eyes that are allowed to heal and develop fully watertight closures after vitrectomy are not expected to face similar inaccuracies.
Finally, the device will benefit from further miniaturization. Although the current device requires a USB connection to a computer to obtain readings, future iterations adapting advancements in wireless technology will enable further miniaturization and portability. Despite these limitations, this device offers a feasible alternative for IOP measurement in patients with altered or artificial corneas.