Intrastromal ring segments (IRS or ICRS) or intrastromal continuous rings (ICR) are small polymeric devices that are introduced in the corneal stroma to regularize the corneal surface and correct high refractive errors. This technique was originally introduced to treat patients suffering from keratoconus (KC),
1 but it was extended to other indications, such as marginal pellucid degenerations,
2 post-LASIK ectasia,
3 and high myopia, in which laser refractive correction is not possible.
4 Three main types of rings are now on the market: continuous rings (e.g., MyoRing, Dioptex GmbH), almost continuous ring segments covering an arc with a central angle between 320° and 355° (e.g., Keraring, Mediphacos, Belo Horizonte, Brazil), and IRS that cover angles below 210° and can be placed by pairs depending on the classification of the cone and amount of correction (e.g., Ferrara, Ophthalmic Ltd.; or Intacs, Addition Technology Inc.). Beside their difference in angular coverage, the existing systems differ by the design of their cross-section and diameter. Nomograms based on morphological parameters, such as the central corneal curvature, minimum corneal thickness,
5,6 or the KC classification,
7 are used to plan the intervention and select the appropriate implant.
Surgical outcomes are characterized by a high variability,
8 thus controlling the postoperative biomechanics is a challenging issue, although several authors
4,5,9,10 reported that the smaller refractive correction owing to the larger diameter is not in contrast with the other observed effects. For example, the shape and size of the cross-section of the implant, its diameter and arc length (in the case of IRS), as well as its implantation position affect the postsurgical shape of the cornea and the respective refractive correction. Also, mechanical factors, such as intraocular pressure (IOP) and tissue biomechanics, play a role and should be accounted for when designing a nomogram.
4,5,7
The implantation depth is one of the surgical parameters that was evaluated clinically. The recommended implantation depth is 75% to 80% of the stromal thickness. Hashemi et al.
11 stressed that for IRS there was an optimal range of stromal depth between 60% and 79% for which the refractive correction was maximal, whereas any other implantation depth had a low impact on the topographic outcomes. Barbara et al.
12 pointed out that the actual insertion depth observed in patients treated with IRS was shallower (∼60%) than the intended insertion depth (∼80%). This observation could be partly explained by the local variation of corneal thickness, whereas the surgical incision remains at a constant distance from the anterior surface but highlighted the difficulty of comparing and interpret existing clinical data. This 20% mismatch in implantation depth or the uncontrolled postsurgical rotation of the implant within the stroma
13 could impact the refractive outcomes.
Unlike the placement of ring segments inside an intrastromal tunnel, continuous rings require a complete intrastromal pocket,
14–16 which has important consequences on the mechanical stability after treatment as the pocket cuts a large surface of the cornea that can include both normal and pathological tissues.
17 Despite providing a higher refractive correction than ring segments, the mechanical impact of the intrastromal pocket remains unclear as it cannot be quantified in clinics. To estimate the mechanical impact of this treatment, Daxer
18 used the Laplace equation to calculate the strengthening corneal factor (SFC), a mechanical marker based on the ratio between the Cauchy stress (σ) in the corneal tissue before and after the ring implantation (SFC = σ
Before / σ
After). Using this simple approach, he estimated an SFC of 2 to 3 for continuous rings and an SFC of ∼1 for ring segments. In his opinion, this difference was explained by the fact that continuous rings restricted the corneal movement acting as an auxiliary limbus, which was not the case for ring segments. Based on this calculation, Daxer suggested that continuous rings were able to introduce a corneal strengthening that should avoid the progression of KC, whereas ring segments would not, as other clinical studies pointed out.
19–21 Nevertheless, other authors outlined the need for additional evidence to confirm this claim.
22
Clinical studies showed that these implants regularized the corneal surface and provided a noticeable correction even for high myopia (>6 diopters [D]). However, planning the surgery to achieve a specific refractive outcome remains challenging.
10 The mechanical principle underlying this treatment remains poorly understood and is difficult to extract only from clinical studies.
8 As clinical studies often present mixed populations with different degrees of myopia, KC severity, or implant typology, it is not possible to isolate the contribution of each parameter on the refractive outcomes. In addition, mechanical properties of the cornea play an important role in the procedure but, unfortunately, it is not possible to characterize corneal biomechanics in vivo with current clinical devices. In silico models have been proposed to study the insertion of intrastromal rings,
9,23–26 however, to the best of our knowledge, no study systematically reported the individual contribution of the implant geometry, the surgical and mechanical parameters on the postsurgical refractive outcomes.
In the present study, we use a calibrated in silico model to understand the mechanical response of the cornea to the treatment and to estimate how different geometric and biomechanical parameters affect the refractive outcomes. Our hypothesis is that implants do not induce a corneal strengthening (change in stresses) but a local mechanical effect, which is a combination of the added volume and the position of the implant with respect to the corneal center, which modifies corneal kinematics and regularizes the corneal surface without introducing a great change in the central stromal stresses. More peripheral implants should have a lower impact on refraction as the localized mechanical effect will dissipate before reaching the corneal center.