Our calculations allow estimation of the impact of an IOL of the same nominal power but differing designs, on the postoperative refraction. For bi or plano-convex implants (shape factors ranged from –1 to +1), the maximum amplitude of displacement of the principal object plane is equal to the central IOL thickness decreased by the distance between the principal planes
\(\overline {{H_i}H{{\rm{^{\prime}}}_i}} \). The impact on postoperative refraction of the ELP
T shift is highly dependent on the optical power of the IOL as suggested by the Gullstrand equation. As the thickness of the implants increases with their power, this tendency is increased, and we calculated that these design variations could induce postoperative refraction variations between approximately 0.50 and 3.0 D, for implant powers ranging from 15 D to 35 D. Therefore, the use of high index, thinner implants reduce the influence of variations in optical design. We did not explore the influence of the corneal geometry and power in the ELP
T prediction. Using paraxial matrix optical calculations, Schröder and Langenbucher
24 have studied the relationship between the thin lens predicted ELP and axial position of a thick IOL achieving the same refraction (respectively referred to as ELP
t and ALP in the present work). They found that the corneal power had less influence than the lens power and design, on the difference between the ELP and the ALP. In all scenarios, the ALP was shorter than the ELP, which was confirmed in the present work where we also found that the ALP of a thick lens must be placed in front of the thin and thick ELP locations for achieving the same refraction. We limited our analysis to biconvex lenses, although these authors included concave–convex minus powered IOLs. Although the latter did increase the discrepancy between ALP and ELP, the influence of variations in the ELP
T distance on postoperative refraction appears to be relatively weak and not clinically significant for low power implants, even for negative implants, whose convex concave design could induce a more marked variation between the position of their vertices and the plane of the ELP
T. The improvement of the calculation formulas for long eyes requires other adjustments than those related to the actual position of the implant, which rather concern the measurement of the posterior segment of the eye and the historical assumptions used by current biometers to infer the axial length from the measured optical path length.
25,26 This finding is reflected in a recent study
27 in which there was no significant difference in the accuracy of thick lens IOL power formula based on calculated versus manufacturer's IOL data for eyes with ALs of 22 mm and more. Fernández et al.
22 suggested modifying the refractive index of the cornea to correct errors beyond the ELP prediction, including assumptions from the biometers.