While the above EZ metrics have been used extensively, there is growing interest in using EZ band intensity or reflectivity as a potentially more sensitive biomarker for evaluating photoreceptor structure, especially in early disease states or following surgical repair of macular hole and retinal detachment (see
Fig. 2E).
20 EZ reflectivity is affected by photoreceptor waveguiding and light scattering,
44 and has been found to be maximally reflective when the OCT beam enters through the pupil center.
45–47 Trauma resulting in commotio retinae often leads to a temporary increase in EZ reflectivity, or EZ disruption, that resolves over time.
48–52 These changes to EZ reflectivity can be monitored, and has been suggested as a biomarker for tracking photoreceptor recovery following clinical intervention.
21,53,54 Conditions with dysfunctional or reduced cones including age-related macular degeneration (AMD),
55,56 RP,
57 and achromatopsia
21,58 often present with reduction in EZ reflectivity, supporting its use as a biomarker for photoreceptor structure and function. Furthermore, reduction in EZ reflectivity has been shown to occur prior to EZ dropout in conditions including nonneovascular AMD and epimacular membrane,
59–61 suggesting that EZ reflectivity might provide a sensitive measure of subclinical deterioration of photoreceptors. Similarly, areas where EZ loss was initially observed in eyes with MacTel type 2 were found to later develop neovascular membranes—these areas must be treated before irreversible damage to photoreceptors occurs.
62 Changes in EZ reflectivity have also been associated with measures of retinal function including visual acuity and retinal sensitivity in conditions such as MacTel type 2,
63 early AMD,
64 Best vitelliform macular dystrophy,
65 and macular hole following surgical repair.
66 Additionally, changes to EZ reflectivity have been correlated to retinal dysfunction and altered blood flow in type 1 diabetes without retinopathy.
67 There are emerging techniques for detection of changes to EZ reflectivity across larger retina areas,
68 including automated methods for quantifying EZ reflectivity.
69,70 Taken together, these findings suggest that quantification of EZ reflectivity has clinical utility for disease detection (both earlier and more accurately) and tracking in a variety of retinal pathologies. Despite growing interest in the use of EZ reflectivity as a potential biomarker of photoreceptor function and structure, multiple challenges to its clinical adoption exist, which we review below.