Table 2 summarizes the measures obtained from traditional subjective refraction and tele-controlled subjective refraction and the statistical comparison between the two methods. Compared to the traditional method, the tele-controlled method did not show statistically significant differences (
P > 0.05) for any parameters. For sphere (S), SE, and cylinder (C) in the total group, the mean differences (traditional subjective refraction – tele-controlled subjective refraction) and their 95% limits of agreement were −0.03 ± 0.42 D, −0.03 ± 0.36 D, and 0.00 ± 0.39 D, respectively. However, even though there were no significant differences between the 2 methods in any measurements, we found wider ranges of the 95% limits of agreement for both J
0 (±0.57 D) and J
45 (±0.79 D). We also analyzed the percentage of refractive parameter differences between the 2 methods that fell within ±0.25 and ±0.50 D (see
Table 2). We observed differences within ±0.25 D in 94% of eyes for SE, in 77% of eyes for J
0 and in 72% of eyes for J
45. As shown in
Table 2, differences in all parameters between the 2 methods within ±0.50 D were ≥84% in all cases.
Figure 2 shows the Bland-Altman plots comparing the SE measured with traditional and tele-controlled methods for the total, young children, teen, and adult groups. As shown in
Figure 2, for SE, the 95% limits of agreement of young children, teens, and adults were ±0.37 D, ±0.41 D, and ±0.33 D, respectively. In addition, we encountered no significant differences (
P > 0.05) for any parameters between eyes in the three age groups (
Table 3). The agreement of SE measured with traditional and tele-controlled methods for the total and each refractive error group is shown in
Figure 3 using Bland-Altman plots.
Table 4 summarizes the statistical comparison and agreement between traditional and tele-controlled subjective refraction for SE, J
0, and J
45 in different refractive error groups. For SE, the mean difference and their 95% limits of agreement of mild hyperopia and emmetropia, mild myopia, and moderate and high myopia groups were 0.04 ± 0.26 D, −0.05 ± 0.38 D, and −0.04 ± 0.36 D, respectively. No significant differences (
P > 0.05) were found between the two methods for any parameters in either refractive error group.