Measurement of the correct amount of astigmatism is important for accurate toric IOL choice. Intra-operative aberrometry during an aphakic state offers the opportunity to implant an IOL by incorporating the total corneal power (anterior and posterior) and the incision induced astigmatism.
2 Whereas ocular surface hydration is a known variable effecting IA, it has not been examined in the past. This study attempted to optimize IA measurements using automation objectively and by subjectively observing the live refractive tracings.
Analysis found DVs for all frame intervals using automated methodology showed no significant difference when compared with IA readings. In 33.3% of cases, IA active refraction traces were stable and the DV (0.27 D ± 0.15 D) was significantly less than unstable traces (0.49 D ± 0.28 D). Alarmingly, 53.3% of unstable runs would have resulted in an unexpected refractive outcome >0.5 D of cylinder had the surgeon gone with the recommendation compared to only 6.7% of stable runs. Currently, this active refraction trace is not available and surgeons are consequently unaware of the potential impact of IOH has on the intra-operative measurement.
In this study, a process of statistical assessment of astigmatism stability was attempted by measuring the average value of a specific frame interval. If the standard deviation of the data was less than 0.5 D during that period, then the astigmatic value was included as a stable. Data were excluded as noise if the standard deviation was more than 0.5 D, as this could lead to a change in toric power. Different intervals were measured; 10 frames to 100 frames and the frames were moved along the trace, hence the moving average. In the ideal world, the perfect signal would have very little variability (i.e. very low standard deviation for the entire run). It would yield the same result and lens recommendation irrespective of frame size and location of the frame on a trace. Furthermore, all the data would be included in the analysis as the standard deviation would be below 0.5 D. However, the ocular surface does not sustain a stable refraction over long intervals with standard deviation below 0.5 D and significant portions of data are excluded as noise with long frames (see
Fig. 4). Both drying of the cornea and perhaps frequent application of BSS likely contribute to these fluctuations.
Figure 4 demonstrates that short intervals, alternatively, have many frames that achieve standard deviation of less than 0.5 D. This, however, fails to account for the general trend and includes more noisy data, leading to an overall higher cylinder prediction. A middle ground, where less data is lost and a tight standard deviation for a reasonable period can be achieved, would be optimal in an automated signal detection system.
Empirically across all generated graphs, 40 frames (88 ms), 50 frames (110 ms), and 60 frames (132 ms) had the best compromise.
Figure 4 shows that around 40 to 60 frames there is a slight “plateau.” Beyond 60 frames there was a significant amount of data loss, and, at less than 40 frames, whereas a lot more data were included, it was likely to be noisier. Difference vectors to the actual postoperative cylinder from these frames’ interval average values were compared. Analysis showed there was no significant difference when comparing IA data provided by the device and the above frame interval analysis. Automation does not appear to help with signal detection in the ever-changing environment. Thereafter, we excluded the frame analysis method and utilized the data directly provided by the device to generate active refractive tracings.
For the first time, by digitizing live data, tracings of live refraction were created postoperatively. A trace was considered stable if the variability was less than 0.5 D, as seen with the naked eye. This threshold was set arbitrarily as 0.5 D or over could result in toric power change. Examining these IA refractive tracings quantitatively, 15 of 45 (33.3%) runs were stable and, in these cases, IA influenced IOL choices for better visual outcomes or confirmed the pre-operative selection (DV = 0.27 D). In the unstable runs, 14 of 30 (46.7%) cases would have resulted in a <0.5 D cylinder outcome, whereas in 16 of 30 (53.3%) of the cases, an unexpected refraction of >0.5 D cylinder would have resulted if the surgeon had gone with the recommendation. In none of these cases, the surgeon would have had access to the live refraction data at the time of IOL choice. Within the unstable runs, in the 5 cases when IA suggested a differing IOL power to the pre-operative choice, the potential resulting DV was 0.59 D, suggesting deferring to the pre-operative choice would yield better outcomes. In the other three cases, the IA suggestion was the same as pre-operative biometry. In these cases, the captures had occurred within the correct range by chance. IA is subject to ocular surface stability in an unblinking eye and its measurement can also vary and mislead the surgeon. The present study highlights the importance that the IOL choice is not an inactive choice from a printout but requires attention toward the quality of measurement irrespective of the source of the measurement.
The issue of toric IOL correction relies on the measurement of the anterior and posterior corneal astigmatism. IA offers the opportunity to measure a single value that includes anterior and posterior astigmatism and does not require a theoretical correction of posterior corneal astigmatism. This has clinical impact in the measurement of low against the rule astigmatism as theoretical formulae will suggest corrections. The low astigmatism value may be due to noise and thus correcting it would needlessly induce the cylinder leading to meridional magnification and spatial distortion should the surgeon act on it.
14 For example, in low toric IOLs, such as T2, can induce astigmatism if non toric IOL is required. The variability in surface stability particularly in the unstable traces clearly highlights the importance of the capture occurring at the correct moment. In addition, presently there is no recommendation from manufacturers regarding the application of BSS to the cornea and the time period between application and measurement depends on the discretion of the surgeon. Incorporating the active refraction tracing into IA devices such that it can be reviewed live would better inform surgeons about the quality and stability of the readings. This is a simple yet useful solution to the current situation where the clinician is unaware of potential fluctuations.
By considering posterior astigmatism, IA can achieve a high rate of refractive outcome predictability especially in patients paying for presbyopic or toric IOLs as well as those with previous refractive surgery (e.g. patients with LASIK/post PRK).
15–19 IA has also been shown to be valuable in the treatment of astigmatism during cataract surgery with both power selection and alignment of the toric IOLs and corneal astigmatic incision titration.
20–24 However, in the literature, there have been varying results of IA accuracy with some studies advocating for caution when interpreting IA measurements with a significant discrepancy compared to pre-operative measurement.
25,26 The range of prediction of postoperative cylinder <0.5 D of the target refraction ranges from 48.1% to 82% across multiple recent studies.
5,15,17,19,27–29 Just as biometry may be affected by dry eyes, fixation, and size of palpebral aperture, a number of factors may confound intra-operative aberrometry measurements, such as IOP, distortion from eyelid speculum, fixation, and vitreous floaters.
2,4,7,30 Studies have demonstrated that by being attentive, operative variables can be adequately controlled during aberrometry measurements and consistent across these studies is the researchers’ optimal efforts to keep surgical conditions constant to minimize confounding error from potential confounding variables (e.g. speculum pressure).
20,22 However, there is no current way to measure ocular surface stability and, as such, without knowledge of the researcher, the surface may be too dry or overhydrated with BSS impacting the measurement. In cases such as these, confounding influences such as variation in IOH may contribute to aberrant readings without the surgeon's knowledge. This study showed 93% of stable runs would have resulted in postoperative refractive outcomes <0.5 D compared to 47% of unstable runs. It is likely that variability in the outcomes of multiple studies may be influenced by a spread of measurements taken at both stable and suboptimal states. Results could be greatly improved if the trace was made visible to the surgeon, reducing prediction error and translation to better outcomes for patients.
IA has been shown to be comparable and superior in instances to formulae in normal and short eyes while providing superior results in long eyes.
5,15,16,20,30,31 With regard to stability of the ocular surface stratified by axial length, there was no bearing. In long eyes, 9 of 12 runs (75%) compared to 2 of 3 (67%) of short eyes and 19 of 30 (63%) of normal eyes had unstable runs. This shows that the axial length likely is unrelated to intraocular surface hydration which effected all lengths of eyes in similar proportions, emphasizing the importance of careful monitoring to obtain the best measurement. With regard to usability and the surgeon learning curve, one survey found 20% of 101 respondents reported it took more than 100 cases to feel comfortable with the ORA.
32 However, 38% reported the transition in the first month, after completing fewer than 30 cases.
32 Including an active trace would expedite this process and avoid incorrect selection of toric lens, which, in low toric cases, can instead induce astigmatism if a non toric was required.
This study has a number of limitations. Exclusion of patients with “abnormal eyes” was based on recorded pre-operative and intra-operative information, which may have omitted minor corneal irregularities or other variables that could have reduced the quality of ORA readings. In a prospective study, these variables could be captured using topography and eyes excluded. Additionally, a single surgeon from a single center performed all surgeries and a limited sample size of 15 patients were included. However, this was mitigated by including 45 individual IA runs and 18,000 data points for power analysis. This study is a pilot proof-of-concept study for future studies that should include greater sample sizes for increased power. Additionally, further research may choose to include the live refractive trace made available to the surgeon at the time of intra-operative aberrometry.
In summary, surgeons should take advantage of the live refraction they see on the screen and ensure it is stable before capturing data. Surgeons should look at the captured refraction and expect it to be similar to the live refraction in sphere, cylinder, and axis. If there is a significant difference, they should consider repeating the measurement.