In this work, we introduced a custom imaging system that has been developed to provide high-resolution, wide depth-of-field, reflection-free, multispectral IR imaging of the Meibomian glands. Our results demonstrate that IR filter of 750 nm is the optimal choice for Meibomian gland imaging because it typically provides images of greatest contrast. Both simple and Michelson contrasts for this wavelength are significantly greater than for other wavelengths (650 nm, 900 nm, 950 nm, and 1000 nm). The wavelength of 1000 nm seems to give the worst contrast results as the values for this wavelength are significantly lower than for other wavelengths, this includes the wavelength of 600 nm, 700 nm, 750 nm, 800 nm, and 850 nm. Furthermore, Meibomian gland contrast (both simple and Michelson) for the wavelength of 750 nm negatively correlates with Meibomian gland depth meaning that glands that are deeper into tarsal plate appear more faded in meibography. Possible reasons for this include an increase in light scatter deeper in the tissue, causing lower contrast. Further investigation may identify imaging enhancements to compensate for or overcome this issue. In addition, as mentioned elsewhere in this article, Meibomian gland depth assessment is not a standard measurement captured with an OCT and is not well-described in the literature; thus, the determination of the apparent Meibomian gland position depended on the subjective assessment of the examiner. This possible correlation requires further investigation in a properly sampled population with different instruments to confirm our findings. Although the wavelength of 750 nm is the optimal choice for best Meibomian gland contrast, our study showed that some individuals may benefit from a different wavelength for imaging based on the overall depth of the tissue. Results presented in
Table 3 show the wide range of wavelengths that works for different people. Those differences imply that Meibomian gland depth may play an important role in meibography. As such, future imaging designs may include range of different wavelengths to optimize image capture.
However, the findings of the current study do not support the previous research. A similar study by Peral et al.
9 suggests that the greatest contrast of Meibomian glands is obtained for wavelength of 600 nm. The authors designed the experimental setup with tunable monochromatic light outputting IR light from 600 to 1050 nm in 25-nm steps. The reported Michelson contrast values ranged from 0.04 to 0.06, similar to our own work (0.05–0.08). In the aforementioned study, the contrast varied across individuals, as was also observed in our study, resulting in the conclusion that Meibomian gland contrast depends on Meibomian gland depth, as shown in the second part of this study. Even though Peral et al.
9 used a more sophisticated method of pixel selection based on local maxima and minima detection, we believe that the images captured at the 600-nm wavelength may provide false results, because other than glands structures become more apparent, such as blood vessels (as seen in
Fig. 4), which seem to be much darker than the actual intergland region; as such, they may contribute to the overall contrast value calculated with Peral et al.
9 method. The authors have reported that the second wavelength that gave the greatest contrast was 775 nm, which would align with the result of the current study.
Keratograph 5M operates at wavelength of 840 nm and LipiView II in an IR spectrum between 890 and 940 nm.
15 This study highlighted that contrast tends to decrease after reaching a peak at 750 nm. This outcome is contrary to that of Peral et al.,
9 who found that Meibomian gland contrast (both simple and Michelson) tends to increase above the 900-nm wavelength. The apparent decrease in Meibomian gland contrast with increasing wavelength observed in our study may relate to the decreased sensitivity of the study camera for longer wavelength IR light. As can be seen in
Figure 4, hot spots (bright areas in the center of the frame) become more apparent for longer wavelengths. Also, hot spots are worse at smaller apertures (larger f numbers), as in our setup. It is possible that the cameras used in commercially available devices have better IR sensitivity and/or IR optics allow them to operate effectively over longer wavelengths.
This experiment provides a new insight into the relationship between Meibomian gland contrast (measured with two different methods) and Meibomian gland depth. Specifically, the current study provides tentative initial evidence that Meibomian gland contrast depends on Meibomian gland depth. These results should be taken into account when imaging Meibomian glands, because glands that are situated deeper into the tarsal plate typically seem to be dimmer on IR photography. Our results show also that those glands may require a longer wavelength for imaging to obtain better contrast. In
Figure 8, a shift in imaging wavelength for deeper situated glands can be seen. This observation is most apparent for the 600-nm wavelength (navy blue line), which gives optimal contrast for shallower glands but much worse for deeper glands. In reviewing the literature, no data were found on the association between Meibomian gland depth and Meibomian gland contrast. To our knowledge, this study is the first to investigate this relationship.
Meibomian gland intensity metrics (metrics based on grayscale level of the image) have been successfully used to track changes of the Meibomian glands.
4,9–11 Yeh and Lin
16 have reported that Meibomian gland contrast (defined as an average difference between the mean intensity along the Meibomian glands and the mean intensity along the regions between the glands, which corresponds with the simple contrast in our study) may be a good diagnostic test for the diagnosis of lipid-deficient dry eye because patients with lipid-deficient dry eye have a significantly lower contrast than controls.
16 Although our study did not investigate the difference between groups, it has shown a high variability in contrast for each wavelength and each participant. The same group have also determined repeatability and limits of agreement of Meibomian gland contrast for Keratograph 5M.
4 It has been shown that contrast changes (in grayscale, 0–255) greater than 11 units in the upper eyelid or 18 units in the lower eyelid are more likely caused by physiological changes rather than the head position or room lighting.
4 Because the flash is much more powerful than the continuous light and has broader IR spectral distribution than typical room lightning, images were independent on ambient lighting in our study. Moreover, the handheld eyelid everter that was adopted to assist in eyelid eversion, provided consistent eyelid eversion across various wavelengths.
A number of studies have proposed and evaluated various Meibomian gland intensity metrics, other than contrast, such as the gland signal index, mean, standard deviation, median, mode, energy, relative energy, entropy, standard deviation irregularity of the selected pixels, kurtosis, and skewness of region of interest histograms.
17,18 It has been shown that those metrics could be a promising biomarker for Meibomian gland dysfunction because patients with a higher level of dropout had significantly lower visibility.
17,18 It has been also shown that those visibility metrics correlate with bulbar redness, tear meniscus height, meibum expressibility score, and noninvasive tear break-up time.
11 Furthermore, it has been proposed that median pixel intensity in combination with ocular surface metrics such as, dropout percentage, tear meniscus height, lid margin abnormality score, and Ocular Surface Disease Index score can be a powerful tool for a Meibomian gland dysfunction diagnosis.
11 Once again, analyses of those studies was based on images obtained with a Keratograph 5M. In our study, we focus on the measurement of contrast, but the image analysis could be easily extended to more parameters. Nonetheless, it is worth highlighting that Meibomian gland reflectivity measurement is of great interest to many researchers and is of significant importance for Meibomian gland health assessment.
Although the study has demonstrated successfully that Meibomian gland contrast varies across individuals and is influenced by Meibomian gland depth, it has certain limitations. The relatively small sample size represents the exploratory nature of this study. Another source of uncertainty is the possibility of measurement errors in manual annotations of Meibomian gland images. Even though the images were labelled by one annotator, it is still a subjective matter to decide whether pixels in the image belong to gland or eyelid region. A more reliable, repeatable, objective method of meibography image segmentation is needed. There are a number of examples in the literature that implement artificial intelligence–based algorithms to perform this task objectively in repeatable and reliable way.
19–25 However, those type of tasks depend on the image type analyzed and the device they have been captured with; as such, it is our future goal to develop an objective method that would be suitable for images captured with the system presented in this work. Another limitation of the current study is that Meibomian gland depth measurements were based on the examiner's judgment of the location of Meibomian glands. A limited number of studies have used OCT to measure depth of Meibomian glands, with little evidence on the measurement of such features with the device.
26,27 Despite its limitations, the study certainly adds to our understanding of the Meibomian gland imaging. Several questions still remain to be answered. A natural progression of this work is to run a larger clinical trial to determine the diagnostic capability of Meibomian gland contrast measured with this newly developed system. More broadly, research is also needed to confirm the Meibomian gland depth measurement with OCT. A further study could also assess Meibomian gland contrast of the upper eyelid glands because they have slightly different morphological characteristics, and it is possible that contrast could differ as well.
An implication of these findings is that both imaging wavelength and Meibomian gland depth should be taken into account when measuring Meibomian gland contrast. Because meibography has been used widely in optometry practice, it is important to better understand its strengths and weaknesses. Many commercially available devices provide processed images, which are good for the subjective analysis of Meibomian gland health. However, image processing often compromises contrast measurement. Thus, we have proposed a new system that is optimized for the assessment of intensity-based Meibomian gland metrics.