The surgical techniques used to remove SiO emulsion
7 include triple air–fluid exchange and BSS lavage,
8 intended to dilute progressively the SiO emulsion. To compare the two, we standardized BSS lavage to a “triple” withdrawal of a volume equal to the volume of the typical vitreous chamber (V
0 = 4 mL).
Previous studies comparing SiO removal techniques were based on the qualitative measure of patients’ floaters,
5 emulsion droplets,
4 or scattering particles in ultrasound video frames,
16 but none quantitatively measured silicon content. Examining the dry residue through XPS allowed unprecedented accuracy, although it also identified the small “background” silicon quota physiologically present in human fluids
17; to discriminate the two fractions, we measured samples of a patient who never received SiO tamponade as a blank reference.
Sample 1 experimental data showed a wide distribution of silicon content across patients of both groups (
Fig. 1), representative of the largely variable “baseline” emulsion at the time of surgery
18 but no difference between groups, as expected. Samples 2 and 3 of the AFX group showed a significantly higher silicon content (
Fig. 2 and
Table 2), and AFX technique allowed the removal of a significantly higher cumulative silicon mass (
Fig. 3 and
Table 3).
Assuming that SiO is removed in the same proportion as fluid (i.e., in the scenario where oil droplets are randomly dispersed in the BSS and the surgical probe collects randomly BSS and oil without any "preference"), both BSSL and AFX should follow the dilution laws of a well-mixed box (
Fig. 4): BSS lavage is a continuous replacement of mixture with clear fluid, whereas AFX consists of two asynchronous phases, a nearly complete removal of the mixture followed by the injection of an equivalent volume of clear fluid. The main difference between BSSL and AFX is the way the solvent is introduced.
Silicon content ratio of consecutive samples is a sensitive indicator of the procedure efficacy, and its value is expected to be constant among consecutive samples in the well-mixed box scenario and can be calculated equal to 1/
e for BSS lavage and (1 – α) for the AFX well-mixed models, respectively (see
Appendix B).
Silicon content ratio of experimental data reported in
Table 4 and
Figure 5 shows that both groups performed much worse than their respective models: if the eye behaved as a neutral container and followed the well-mixed box model, in fact, more effective procedures would give lower successive sample ratios.
The rapid decrease of the SiO content, given by the α = 0.90 procedure (
Fig. 4), corresponds to a ratio of the different samples as small as 0.1, much lower than the ratio (0.5) resulting from the α = 0.50 procedure; in the same scenario, BSSL, which in
Figure 4 exhibits the slower decrease of SiO content, would give the lower ratio: 0.37 (corresponding to 1/
e). Real-life emulsion behaves much differently since our experimental data show an average silicon ratio between consecutive samples of 0.58 for BSS lavage and 0.9 for AFX (
Table 4,
Fig. 5), both much higher than expected in a well-mixed box, meaning that both assumptions of silicon removal in the same proportion as fluid and the eye wall as a neutral container are false and must be rejected.
There are two main reasons explaining this macroscopic deviation from the well-mixed box model: the nonneutral role of the eye wall in retaining SiO and the active search by the surgeon of SiO emulsion floating over the fluid–air interface during AFX. In principle, the surgeon actively seeking SiO emulsion should have a higher mass removal compared to the box model, whereas the emulsion adherence to the eyewall should decrease the effectiveness of the procedure.
Figure 5, comparing expected and observed data, clearly suggests that the second effect dominates and the eye wall “actively retains” silicon: the vitreous chamber is not a “neutral” container.
Therefore, our data indicate that the silicon content of all examined samples represents a fraction of the total intraocular silicon mass present in the eye, and the “eye wall” is far from being neutral but behaves as a “buffer” dynamically exchanging SiO emulsion with its liquid content according to complex (and still unknown) mechanisms.
It should be noted that the macroscopic SiO bubble and its emulsion contact several anatomic structures, including the optic nerve, retina, pars plana, pars plicata, ciliary body, posterior iris, Zinn's zonule, and crystalline lens posterior capsule, whose intricate anatomy, roughness, polarity, and charge distribution can hide, enclose, or bind SiO emulsion.
The relative efficiency of both methods compared to their respective theoretical model is shown by the distance between experimental data and the model in
Figure 5. Why so little silicon is removed during the triple AFX compared to its theoretical model (
Fig. 4 and
Fig. 5) remains puzzling: emulsified SiO adheres to biologic tissues
1; when fluid is exchanged to air, the increase in surface tension may favor droplet adherence to rougher surfaces such as the pars plicata and iris. Similarly, the supernatant SiO at the fluid–air interface may “seed” the emulsion over the eye walls as fluid is removed just like seashells left on the shore. This may explain the mechanism of the SiO “buffer” and the presence of silicon in comparable concentrations throughout the subsequent samples discussed in detail in
Appendix B. It should also be noted that the roughness of the inner vitreous chamber, especially at the iris and pars plicata, makes its surface incomparably wider than a sphere portion.
Shiihara et al.
16 concluded that AFX decreased its effectiveness compared to BSSL as the axial length increased: a notion consistent with the above argument since longer axial length determines a wider surface. Yu et al.
8 compared the number and size of droplets after AFX and BSSL using a Coulter counter and reported a 35% to 40% of reduction in SiO bubbles between 1 and 12 microns. This figure does match our results. However, it should be noticed that Yu et al.
8 considered only a range of droplet sizes, whereas XPS allows a much more accurate detection of SiO.
4 As a matter of fact, the total SiO content is related not only to the droplet size but also to their number. Therefore, it is not possible to assess the SiO contained in the droplets not considered in their count.
It is worth noting that we do not know the overall silicon content at time 0 (m0), nor we can reliably estimate it. However, the ratio of successive samples (Δm3/Δm2 and Δm2/Δm1) that we measure is quite high, compatible with a very low overall silicon removal efficacy (Δm/m0; i.e., the ratio of the diminution of silicon mass to the initial mass) of both techniques, presumably less than 10% to 20%. This notion suggests that regardless of the technique used, the surgeon should probably aim at introducing a solvent volume at least five to six times the vitreous chamber.
Although AFX removed more silicon than lavage, it proved less efficient than BSSL when compared to its respective theoretical model. This may be clinically relevant since it is probably easier and less surgically dangerous to prolong BSS lavage than to perform multiple air–fluid exchanges in order to increase the fraction of removed SiO (
Fig. 6).
It is also conceivable that the infusion cannula laminar “jet flow” impinging the retina during BSS lavage may favor the detachment of SiO emulsion adherent to it and its successive aspiration, while the air bubble generated during the air exchange may result in laying a uniform coating of SiO over the vitreous chamber walls. If this is proved correct, a directional cannula held by the surgeon and “sweeping” the vitreous chamber walls may increase the effectiveness of the procedure.
In summary, for the first time, we used XPS as an objective and quantitative measure of silicon content in the subsequent intraocular fluid samples collected during SiO emulsion removal with two different surgical techniques. AFX removed a higher mass of silicon across the three lavages compared to BSS lavage, but none behaved as expected unless we postulate that the eye walls have a high “intrinsic” capability of interaction and retainment of SiO during the procedure.
Our data suggest that SiO dispersion establishes a complex relationship with the eye wall, with biologic, chemical, and electrostatic properties playing an important role in the removal dynamics.
The pitfalls of present study include the relatively small sample size and the novelty of XPS for the present purposes that find very few, if any, comparable studies in the literature. Also, it should be noted that 1000 cS SiO was used for all our patients, and therefore conclusions may not apply to different viscosity compounds. On the other hand, XPS is a sensitive method extensively used for other purposes, and each measure is based on readings from 10 droplets of the same sample to increase data robustness.