Open Access
Artificial Intelligence  |   March 2025
Retinal Microvascular Biomarker Assessment With Automated Algorithm and Semiautomated Software in the Montrachet Dataset
Author Affiliations & Notes
  • Pétra Eid
    Ophthalmology Department, Dijon University Hospital, Dijon, France
    Centre des Sciences du Goût et de l'Alimentation, AgroSup Dijon, CNRS, INRAE, Université Bourgogne, Dijon, France
  • Abderrahmane Bourredjem
    CIC 1432, Epidémiologie Clinique, Centre Hospitalier Universitaire Dijon-Bourgogne, Dijon, France
  • Atif Anwer
    Institut de Chimie Moléculaire Université de Bourgogne (ICMUB), Imagerie Fonctionnelle et moléculaire et Traitement des Images Médicales (IFTIM), Burgundy University, Dijon, France
  • Catherine Creuzot-Garcher
    Ophthalmology Department, Dijon University Hospital, Dijon, France
    Centre des Sciences du Goût et de l'Alimentation, AgroSup Dijon, CNRS, INRAE, Université Bourgogne, Dijon, France
  • Pearse Andrew Keane
    NIHR Biomedical Research Centre at Moorfields Eye Hospital NHS Foundation Trust, London, UK
    Institute of Ophthalmology, University College London, London, UK
  • Yukun Zhou
    NIHR Biomedical Research Centre at Moorfields Eye Hospital NHS Foundation Trust, London, UK
    Institute of Ophthalmology, University College London, London, UK
  • Siegfried Wagner
    NIHR Biomedical Research Centre at Moorfields Eye Hospital NHS Foundation Trust, London, UK
    Institute of Ophthalmology, University College London, London, UK
  • Fabrice Meriaudeau
    Institut de Chimie Moléculaire Université de Bourgogne (ICMUB), Imagerie Fonctionnelle et moléculaire et Traitement des Images Médicales (IFTIM), Burgundy University, Dijon, France
  • Louis Arnould
    Ophthalmology Department, Dijon University Hospital, Dijon, France
    Pathophysiology and Epidemiology of Cerebro-Cardiovascular Diseases (PEC2), (EA 7460), Faculty of Health Sciences, Université de Bourgogne, Dijon, France
  • Correspondence: Louis Arnould, Ophthalmology Department, University Hospital, 14 rue Paul Gaffarel, Dijon Cedex 21079, France. e-mail: [email protected] 
Translational Vision Science & Technology March 2025, Vol.14, 13. doi:https://doi.org/10.1167/tvst.14.3.13
  • Views
  • PDF
  • Share
  • Tools
    • Alerts
      ×
      This feature is available to authenticated users only.
      Sign In or Create an Account ×
    • Get Citation

      Pétra Eid, Abderrahmane Bourredjem, Atif Anwer, Catherine Creuzot-Garcher, Pearse Andrew Keane, Yukun Zhou, Siegfried Wagner, Fabrice Meriaudeau, Louis Arnould; Retinal Microvascular Biomarker Assessment With Automated Algorithm and Semiautomated Software in the Montrachet Dataset. Trans. Vis. Sci. Tech. 2025;14(3):13. https://doi.org/10.1167/tvst.14.3.13.

      Download citation file:


      © ARVO (1962-2015); The Authors (2016-present)

      ×
  • Supplements
Abstract

Purpose: To compare automated and semiautomated methods for the measurement of retinal microvascular biomarkers: the automated retinal vascular morphology (AutoMorph) algorithm and the Singapore “I” Vessel Assessment (SIVA) software.

Methods: Analysis of retinal fundus photographs centered on optic discs from the population-based Montrachet Study of adults aged 75 years and older. Comparison and agreement evaluation with intraclass correlation coefficients (ICCs) between SIVA and AutoMorph measures of the central retinal venular and arteriolar equivalent, arteriolar–venular ratio, and fractal dimension.

Results: Overall, 1069 fundus photographs were included in this study. The mean age of the patients was 80.04 ± 3.94 years. After the image quality grading process with an optimal threshold, the lowest rejection rate was 51.17% for the AutoMorph analysis (n = 522). The measure of agreement between SIVA and AutoMorph retinal microvascular biomarkers showed a good correlation for vascular complexity (ICC, 0.77–0.47), a poor correlation for vascular calibers (ICC, 0.36–0.23), and no correlation for vascular tortuosity. Significant associations between retinal biomarkers and systemic variables (age, history of stroke, and systolic blood pressure) were consistent between SIVA and AutoMorph.

Conclusions: In this dataset, AutoMorph presented a substantial rejection rate. SIVA and AutoMorph provided well-correlated measurements of vascular complexity and caliber with consistent clinical associations. Further comparisons are needed before a transition is made from semiautomated to automated algorithms for the analysis of retinal microvascular biomarkers.

Translational Relevance: Open source software needs to be compared with former semiautomated software for retinal microvascular biomarkers assessment before transition in daily clinic and collaborative research.

Introduction
Quantitative analysis of retinal microvascular biomarkers based on retinal fundus photographs has been extensively performed since the early 2000s. Semiautomated software applications such as Singapore “I” Vessel Assessment (SIVA; National University of Singapore, Singapore), Integrative Vessel Analysis (IVAN; University of Wisconsin, Madison), and Vascular Assessment and Measurement Platform for Images of the Retina (VAMPIRE; VAMPIRE group)13 can provide a thorough description of the retinal vascular network with geometric parameters (i.e., arteriolar and venular caliber, tortuosity, and fractal dimension [FD]).47 These quantitative retinal microvascular biomarkers were shown to be associated with a long-term risk of mortality and ischemic stroke (narrower arterioles and wider venules),810 coronary heart mortality (impaired FD),11 cardiovascular risk factors,1214 and an increased cardiovascular risk profile (suboptimal retinal vascular network).15 Thus, retinal microvascular biomarkers based on fundus photographs could provide insights into vascular systemic disease. This field of research can be described as oculomics.16 
These semiautomated software applications require human, trained operators and involve time-consuming interventions for vessel segmentation correction and for artery–vein and optic disc identification, which increases human workload and decreases work reproducibility. To lessen these flaws and to increase the dataset volume, fully automated applications have emerged recently, with new artificial intelligence image-processing software.1721 Numerous feature-based algorithms have been developed in recent years, such as SIVA DL, Deep Approximation of Retinal Traits, QUantitative Analysis of Retinal vessel Topology and size (QUARTZ), and LUNet.2225 These deep learning approaches aim to decrease the time required for image analysis and enable the exploration of large fundus datasets. Recently, Zhou et al.26 developed an open source deep learning algorithm providing automated analysis of retinal vascular morphology from retinal fundus photographs (automated retinal vascular morphology [AutoMorph]). 
Several studies have explored the agreement between semiautomated software and reported inconsistent results.19,21 However, few data are available on the comparison between semiautomated and fully automated software used to extract retinal microvascular biomarker data.27 These investigations are mandatory for the generalizability of the results and for fundus-based pooled data analysis. To the best of our knowledge, very few data are available on the agreement between SIVA and AutoMorph measurements of retinal microvascular features. 
The current study aimed to compare SIVA software and AutoMorph measurements of retinal microvascular biomarkers in a population-based study of adults aged 75 years and older. The secondary objective was to elucidate whether associations between SIVA retinal microvascular biomarkers and systemic variables in the Montrachet Study were consistent when measured by AutoMorph. 
Methods
Study Population
The Montrachet (Maculopathy Optic Nerve nuTRition neurovAsCular and HEarT diseases) Study is a French population-based study designed to assess the associations between age-related eye diseases and neurological and heart diseases in older participants.28 Briefly, the Three-City Study (3C Study) included 9294 non-[institutionalized individuals aged 65 years and older who were randomly selected from the electoral rolls of three French cities (Bordeaux, Dijon [n = 4931], and Montpellier). Baseline examinations (January 1999 to March 2001) consisted of a face-to-face interview (recording sociodemographic and lifestyle characteristics as well as medical history), a physical and cognitive examination, and fasting blood sampling. 
The Montrachet Study, described in detail elsewhere,29 consisted of a comprehensive eye examination (including fundus photography) in the Department of Ophthalmology of Dijon University Hospital, proposed to participants of the 10-year follow-up of 3C-Dijon study. Overall, 1153 participants agreed to take part and were recruited. All 3C and Montrachet participants gave their written informed consent; the ethics committee of the University Hospital of Kremlin-Bicêtre and the local university hospital ethics committee, respectively, approved the study protocols. The present study adhered to the tenets of the Declaration of Helsinki and was registered as 2009-A00448–49. 
To be eligible for the present work, individuals had to be participants in the Montrachet Study who had fundus images with measurements of retinal microvascular biomarkers via SIVA software and the AutoMorph algorithm. We excluded participants who had the following characteristics: epiretinal membrane and an ophthalmological condition interfering with the macular or peripapillary retinal vasculature. 
Retinal Fundus Photographs and SIVA Postprocessing
All patients had retinal fundus photographs taken after pupil dilation with tropicamide 0.5% (Thea, Clermont-Ferrand, France), centered on the optic disc, using a fundus camera (TRC NW6S, Topcon, Tokyo, Japan). For each participant, one eye was retained for analysis based on the following criteria: right eye for participants born in even-numbered years and left eye for participants born in odd-numbered years; for single-eye patients, the functional eye was selected; and if retinal fundus photographs were uninterpretable for one eye, the other one was retained.15 Retinal fundus photographs were postprocessed using SIVA. Fundus photographs were anonymously sent to the reading center in Yamagata University, Japan. A single trained grader extracted retinal vessel characteristics using SIVA. A fundus photograph was considered uninterpretable if blurred, if zone C could not be analyzed or if the fundus photograph contained fewer than six arterioles and veins. To study the retinal vascular network, the analysis was based on the six largest arterioles and veins from the optic disc center to three successive zones corresponding to 0.5 optic disc diameter to the margin of the optic disc (zone A), 1 optic disc diameter (zone B), and two optic disc diameter (zone C) (Supplementary Fig. S1).2 First, geometric retinal vascular features including FD, vascular caliber, tortuosity, and branching angle were collected and measured by SIVA. An algorithm automatically identified the optic disc and automatically extracted vascular structure and tracing. Then, retinal arterioles (drawn in red) and venules (drawn in blue) were identified and drawn. At the end, algorithms were used to compute quantifiable measurements of retinal vasculature. Human manual intervention consisted of correcting the vessel drawings and removing artifacts. 
Vascular Morphology Feature Measurement With AutoMorph
The same retinal fundus photographs were also analyzed by means of the AutoMorph algorithm, a publicly available and open source algorithm (source code download available at https://github.com/rmaphoh/AutoMorph).26 The analysis was based on the same standardized zones as SIVA, corresponding with the annulus from 0.5 to 1.0 optic disc diameter from the optic disc margin (zone B) and to the annulus from 0.5 to 2.0 optic disc diameter (zone C). 
Statistical Analyses and Method Comparison
Retinal parameters that appeared to be analyzed by both applications were compared in this study. The features measured by both applications were as follows: the central retinal venular equivalent (CRVE), the central retinal arteriolar equivalent (CRAE), the arteriolar–venular ratio (AVR), and the vessel complexity measured by FD. The vessel tortuosity was provided by distance measurement tortuosity in AutoMorph and simple tortuosity in SIVA. 
As a first step in the process, retinal fundus photographs were classified as gradable or ungradable using AutoMorph, according to a grading quality threshold. This threshold of interpretable images can be set from 0 to 1. 
Statistical analyses were conducted using SAS software version 9.4 (SAS institute Inc., Cary, NC). Statistical significance was set at a P value of less than 0.05. Continuous variables are expressed as medians (interquartile range) or means ± standard deviation according to their distribution. Categorical variables are expressed as numbers (percentage). The agreement between the two methods was assessed graphically using scatterplots and Bland–Altman plots and quantitatively using type 3 intraclass correlation coefficients (ICCs) with their 95% confidence intervals (CIs).30 Strength of agreement was classified using the categories suggested by Hahn et al.31 Spearman's correlations assessed the external consistency of the two methods used for analysis of the fundus photographs in terms of the strength of associations with systemic variables (age, sex, treatment for systemic hypertension, body mass index, hypoglycemic treatment, cholesterol-lowering treatment, history of cardiovascular disease and stroke, HeartScore, and major adverse cardiac and cerebrovascular events (MACCE score). History of cardiovascular disease, stroke, and MACCE were variables self-reported by the participants. 
This study adhered to the Guidelines for Reporting Reliability and Agreement Studies (GRRAS).32 
Results
Overall, 1069 fundus photographs were gradable with SIVA software in the Montrachet Study (Fig. 1). The AutoMorph algorithm analyzed 315 fundus photographs, with a threshold of 0.25. With sequential modification of the threshold, 522 fundus photographs were graded using a threshold of 0.50, 0.75, and 1.00. The lowest rejection rate was 51.17%. The images with uneven illumination and blurry conditions were rejected by AutoMorph. The mean age of the participants analyzed with both SIVA and AutoMorph was 80.04 ± 3.94 years and 69% were female. Table 1 presents the characteristics of the Montrachet Study participants at their inclusion in the study compared with participants with AutoMorph analysis (n = 522) and participants without AutoMorph analysis (n = 547). There was no statistically significant differences between the two groups, except in sex, mean systolic blood pressure, and rate of stroke. 
Figure 1.
 
Flow chart of the method comparison study between Montrachet SIVA and AutoMorph.
Figure 1.
 
Flow chart of the method comparison study between Montrachet SIVA and AutoMorph.
Table 1.
 
Characteristics of Participants With AutoMorph Analysis at Their Inclusion in the Montrachet Study (n) Compared With Participants With no AutoMorph Analysis (n′)
Table 1.
 
Characteristics of Participants With AutoMorph Analysis at Their Inclusion in the Montrachet Study (n) Compared With Participants With no AutoMorph Analysis (n′)
The strength of agreement between SIVA and AutoMorph analyses (Table 2) was moderate to good for the FD of arterioles (ICC, 0.47; 95% CI, 0.40–0.53), of veins (ICC, 0.53; 95% CI, 0.46–0.59), and of all vessels (ICC, 0.77; 95% CI, 0.73–0.80). The strength of agreement was poor to slight for all retinal caliber parameters (for CRAE: ICC, 0.23; 95% CI, 0.14–0.32] in zone B and ICC, 0.21; 95% CI, 0.13–0.30] in zone C; for CRVE: ICC, 0.34; 95% CI, 0.26–0.42] in zone B and ICC, 0.34; 95% CI, 0.26–0.41] in zone C). For AVR, there was also a poor to slight agreement in zone B (ICC, 0.32; 95% CI, 0.24–0.40) and in zone C (ICC, 0.36; 95% CI, 0.29–0.44). Figure 2 presents the Bland–Altman and scatter plots illustrating the relationship between AutoMorph and SIVA measurements for these vascular retinal parameters in zone C. Concentration around the first bisector was better for CRVE than for CRAE. For AVR, AutoMorph tended to give higher ratios than SIVA (by a mean estimation of 0.07 units in range of 0.5–1.0 units), but 95% of the differences were within the precision band. For FD, AutoMorph tended to give systematically lower values than SIVA (by a mean estimation of 0.13 units in a global range of 0.83–1.54 units), with a strong proportionality link and 96% of differences within the precision band. However, there was no agreement for vascular tortuosity parameters between the SIVA and AutoMorph analyses, either for arterioles and for veins (ICC, 0.0022 [95% CI, 0–0.09] for arterioles and ICC, 0.0008 [95% CI, 0–0.09] for veins). There was no proportionality link, a mean difference of −2.5 points in a disproportional range, and more than 40% of the differences were out of the precision band (figures not shown). Figure 3 presents examples of fundus photographs with good, moderate, and poor concordance for FD between AutoMorph and SIVA. 
Table 2.
 
Retinal Vascular Parameter Agreement Between SIVA and AutoMorph
Table 2.
 
Retinal Vascular Parameter Agreement Between SIVA and AutoMorph
Figure 2.
 
Bland–Altman and scatter plots illustrating correlation of retinal vessel parameter measurements in zone C between SIVA and AutoMorph.
Figure 2.
 
Bland–Altman and scatter plots illustrating correlation of retinal vessel parameter measurements in zone C between SIVA and AutoMorph.
Figure 3.
 
Fundus photography examples illustrating concordance in FD between SIVA and AutoMorph.
Figure 3.
 
Fundus photography examples illustrating concordance in FD between SIVA and AutoMorph.
Spearman correlation analyses were performed to assess whether the associations between SIVA software retinal features and systemic conditions in the Montrachet Study were consistent when measured with AutoMorph (Table 3). Most associations between retinal vascular parameters found with SIVA were also found with AutoMorph, such as for age and CRAE (R1 = −0.26; P < 0.0001 and R2 = −0.14; P = 0.001), for age and CRVE (R1 = −0.18; P < 0.0001 and R2 = −0.12; P = 0.007), for history of stroke and CRVE (R1 = −0.11; P = 0.013 and R2 = −0.092; P = 0.037), for systolic blood pressure and AVR (R1 = −0.129; P = 0.0003 and R2 = −0.13; P = 0.002), and for history of stroke and AVR (R1 = 0.089; P = 0.045 and R2 = 0.12; P = 0.007). We also found a notably statistically significant association, but only with AutoMorph, for body mass index and CRAE (R1 = 0.14; P = 0.002) and for body mass index and CRVE (R1 = 0.14; P = 0.002). In contrast, we found a statistically significant association, but only with SIVA, for systolic blood pressure and CRAE (R2 = −0.139; P = 0.001), for body mass index and AVR (R2 = −0.10; P = 0.019), for HeartScore and CRVE (R2 = 0.10; P = 0.021), and for HeartScore and AVR (R2 = −0.14; P = 0.001). 
Table 3.
 
Association Between Retinal Vascular Parameters Measured With AutoMorph and SIVA and Systemic Variables
Table 3.
 
Association Between Retinal Vascular Parameters Measured With AutoMorph and SIVA and Systemic Variables
Discussion
In this study, we conducted an in-depth comparison between two different applications to measure retinal microvascular biomarkers from retinal fundus photographs in a dataset of older adults. The SIVA application, a semiautomated software, enabled the analysis of more images than AutoMorph, which rejected 51.17% of the fundus photographs after the image quality grading process. In this specific dataset, AutoMorph seemed to be more sensitive to image quality, excluding a large proportion of retinal fundus photographs analyzed by SIVA. With SIVA, the quality analysis and rate of rejection are determined by manual human intervention, whereas in AutoMorph this step is automated. With AutoMorph, the threshold of interpretable images can be modified from 0 to 1. In this study, various thresholds were tested in a prior analysis to choose the lowest possible threshold for obtaining the highest number of analyzed images. It should be mentioned that there was no improvement in the rejection rate above the 0.50 threshold in this dataset. 
In a study based on UK Biobank data analyzed with automated quantification of retinal vessel morphometry using QUARTZ software, 28.4% of the images were labeled as inadequate.33 The study confirmed the high rate of uninterpretable images when using automated analysis algorithms. It remains important to know the target population characteristics of the dataset used to train and validate the algorithm, because this factor could explain a high rejection rate in another target population. 
Interestingly, in the study based on UK Biobank data analyzed with VAMPIRE, comparing groups with and without analyzable photographs showed that advancing age was associated with a decreased proportion of patients with analyzable photographs, whereas cardiovascular diseases or cardiovascular risk factors did not have a major influence on image quality.34 Our results were not in line with these findings, because older age was not statistically significant in our analysis. 
AutoMorph and SIVA had a moderate to good agreement regarding vascular complexity evaluated by FD and a poor to slight agreement regarding vascular caliber (CRAE, CRVE, and AVR). However, vascular tortuosity was not well-correlated between AutoMorph and SIVA. We could, therefore, assume that the tortuosity measurement methods were different between the two algorithms. Indeed, in AutoMorph the vascular tortuosity was measured in three different ways—distance tortuosity, squared curvature tortuosity, and tortuosity density—whereas in SIVA only one method was provided with curvature tortuosity.3,9,35,36 More harmonized measurement protocols for geometric vascular biomarker could help to reach better agreement. 
Moreover, there were discrepancies between the Automorph and Montrachet datasets that could explain such poor agreement. Indeed, the Automorph algorithm was tested and validated in datasets with younger participants.3745 Fundus photograph device specificity, image compression, image format, or media opacity (lens status) could also explain part of this difference. Table 4 summarizes the population characteristics that could potentially influence retinal microvascular biomarker analysis. Further studies should investigate at which level the heterogenicity of images impact automatic software for a chosen target population. 
Table 4.
 
Characteristics of Patients in the Montrachet Cohort and Datasets Used for AutoMorph Training and Validation
Table 4.
 
Characteristics of Patients in the Montrachet Cohort and Datasets Used for AutoMorph Training and Validation
Based on a similar methodology to our work, a recent study focused on comparing IVAN, a semiautomated algorithm, and the Retina-based Microvascular Health Assessment System (RMHAS), an automated deep learning algorithm.27 The ICCs between IVAN and RMHAS were higher than in our study, ranging from 0.42 for AVR to 0.76 for CRVE. This result could be explained by the different age groups included. The mean age in the study comparing IVAN and RHMAS was 48.88 ± 8.38 years, whereas participants in our study were older, with a mean age of 80.04 ± 3.94 years. Thus, we could assume that there were more retinal fundus photographs with interfering retinal anomalies or with more artifacts in the Montrachet dataset, which can lead to a more difficult segmentation and a lower ICC. As presented in Figure 3, concordance between SIVA and AutoMorph seemed to decrease as image quality deteriorated. 
Associations of age, history of stroke, and systolic blood pressure with retinal microvascular biomarkers remained consistent with both SIVA and AutoMorph. This point is crucial for future pooled dataset analyses based on fundus photographs. Similar comparisons are needed to confirm that the associations between retinal microvascular biomarkers and systemic vascular features are independent of the fundus image analysis strategy (semiautomated or fully automated). Moreover, the comparisons are also needed to confirm that the transition from semiautomated to fully automated algorithms is coherent with previous studies. 
A recent study showed poor stability between intervisit and intravisit retinal parameter measurements by AutoMorph from fundus photographs.46 This instability was even more marked for parameters derived from retinal vascular segmentation, such as vascular tortuosity and FD, than for those derived from optic nerve segmentation. In our study, patients were not retested on several fundus photographs to compare the stability of the parameters studied. It is conceivable that image quality may influence these parameters, which was found to have a moderate effect in the study by Giesser et al.46 The path to validating retinal parameters as biomarkers of systemic pathologies will require stabilization of these parameters. 
The automated feature could make AutoMorph more suitable for analysis of big datasets compared with SIVA. In addition, human involvement in vessel segmentation and identification of the arterioles, veins, and optic disc can lead to potential errors and differences in interpretation if various operators are involved in the data analysis. 
This study provided further external validation for the automated analysis of retinal fundus photographs by AutoMorph. 
We acknowledge some limitations to this study. First, a small number of fundus images was analyzed compared with other larger studies. Second, SIVA Deep Learning analysis could improve our understanding of limited agreement between semiautomated and automated algorithms. Third, this study was based on a Caucasian population aged more than 75 years, which limits the generalizability of these results. 
Further method comparison studies are needed to compare the technical aspects of these imaging software with a larger dataset of sufficient image quality. 
Conclusions
In this study comparing SIVA and AutoMorph, we found a moderate to good correlation for vascular complexity analysis and a poor to slight correlation for vascular caliber analysis, but no agreement on the tortuosity parameters, with a substantial image rejection rate. Although AutoMorph presented promising results in the fully automated analysis of retinal photographs, further research is needed to promote interchangeability and pooled analyses using this software in a large-scale population-based study. 
Acknowledgments
The authors thank Christine Binquet, the CIC team and Laure-Anne Steinberg for her precious help in managing Montrachet dataset. 
Supported by a UK Research & Innovation Future Leaders Fellowship (MR/T019050/1 to PAK) and The Rubin Foundation Charitable Trust. We acknowledge Dorine Maillet for her initial support in this project. 
The Montrachet study was funded by an inter-regional grant (PHRC), the Regional Council of Burgundy, INRA, CNRS, Université de Bourgogne, Regional Council of Burgundy France (PARI Agrale 1), FEDER (European Funding for Regional Economic Development), the VISIO Foundation for children and visually impaired adults, and a French Government grant managed by the ANR under the “Investment for the Future” Program (ANR-11-LABX-0021-01-LipSTIC Labex), Fondation Avenir. 
Disclosure: P. Eid, None; A. Bourredjem, None; A. Anwer, None; C. Creuzot-Garcher, Novartis (C, F, R), Abbvie (C, R), Bayer (C, R), Roche (C), Horus (C), Thea (C); P.A. Keane, America (C), Topcon (C, R, S), Roche (C), Boehringer-Ingelheim (C, S), Bitfount (C), Zeiss (C), Novartis (C, S), Gyroscope (C), Apellis (C), Abbvie (C), Topcon (R, S), Bayer (S), Boehringer-Ingleheim, RetinAI (S); Y. Zhou, None; S. Wagner, None; F. Meriaudeau, None; L. Arnould, Thea (C, R), Horus (R) 
References
Cheung CY, Hsu W, Lee ML, et al. A new method to measure peripheral retinal vascular caliber over an extended area. Microcirculation. 2010; 17: 495–503. [PubMed]
Cheung CY, Tay WT, Mitchell P, et al. Quantitative and qualitative retinal microvascular characteristics and blood pressure. J Hypertens. 2011; 29: 1380–1391. [CrossRef] [PubMed]
Cheung CY, Zheng Y, Hsu W, et al. Retinal vascular tortuosity, blood pressure, and cardiovascular risk factors. Ophthalmology. 2011; 118: 812–818. [CrossRef] [PubMed]
Knudtson MD, Lee KE, Hubbard LD, et al. Revised formulas for summarizing retinal vessel diameters. Curr Eye Res. 2003; 27: 143–149. [CrossRef] [PubMed]
Sasongko MB, Wong TY, Nguyen TT, et al. Retinal vascular tortuosity in persons with diabetes and diabetic retinopathy. Diabetologia. 2011; 54: 2409–2416. [CrossRef] [PubMed]
Sasongko MB, Hodgson LA, Wong TY, et al. Correlation and reproducibility of retinal vascular geometric measurements for stereoscopic retinal images of the same eyes. Ophthalmic Epidemiol. 2012; 19: 322–327. [CrossRef] [PubMed]
Cheung CY, Thomas GN, Tay W, et al. Retinal vascular fractal dimension and its relationship with cardiovascular and ocular risk factors. Am J Ophthalmol. 2012; 154: 663–674.e1. [CrossRef] [PubMed]
Seidelmann SB, Claggett B, Bravo PE, et al. Retinal vessel calibers in predicting long-term cardiovascular outcomes: the Atherosclerosis Risk in Communities Study. Circulation. 2016; 134: 1328–1338. [CrossRef] [PubMed]
Witt N, Wong TY, Hughes AD, et al. Abnormalities of retinal microvascular structure and risk of mortality from ischemic heart disease and stroke. Hypertension. 2006; 47: 975–981. [CrossRef] [PubMed]
Kawasaki R, Xie J, Cheung N, et al. Retinal microvascular signs and risk of stroke: the Multi-Ethnic Study of Atherosclerosis (MESA). Stroke. 2012; 43: 3245–3251. [CrossRef] [PubMed]
Liew G, Mitchell P, Rochtchina E, et al. Fractal analysis of retinal microvasculature and coronary heart disease mortality. Eur Heart J. 2011; 32: 422–429. [CrossRef] [PubMed]
Chew SK, Xie J and Wang JJ. Retinal arteriolar diameter and the prevalence and incidence of hypertension: a systematic review and meta-analysis of their association. Curr Hypertens Rep. 2012; 14: 144–151. [CrossRef] [PubMed]
Daien V, Carriere I, Kawasaki R, et al. Malnutrition and retinal vascular caliber in the elderly: the POLA study. Invest Ophthalmol Vis Sci. 2014; 55: 4042–4029. [CrossRef] [PubMed]
Kifley A, Liew G, Wang JJ, et al. Long-term effects of smoking on retinal microvascular caliber. Am J Epidemiol. 2007; 166: 1288–1297. [CrossRef] [PubMed]
Arnould L, Binquet C, Guenancia C, et al. Association between the retinal vascular network with Singapore “I” Vessel Assessment (SIVA) software, cardiovascular history and risk factors in the elderly: the Montrachet study, population-based study. PLoS One. 2018; 13: e0194694. [CrossRef] [PubMed]
Wagner SK, Fu DJ, Faes L, et al. Insights into systemic disease through retinal imaging-based oculomics. Transl Vis Sci Technol. 2020; 9: 6. [CrossRef] [PubMed]
Mirsharif Q, Tajeripour F and Pourreza H. Automated characterization of blood vessels as arteries and veins in retinal images. Comput Med Imaging Graph. 2013; 37: 607–617. [CrossRef] [PubMed]
Welikala RA FM, Habib MM, Daniel-Tong S, Yates M, Foster PJ, et al. Automated quantification of retinal vessel morphometry in the UK biobank cohort. 2017 Seventh International Conference on Image Processing Theory, Tools and Applications (IPTA). 2017;1–6.
Yip W, Tham YC, Hsu W, et al. Comparison of common retinal vessel caliber measurement software and a conversion algorithm. Transl Vis Sci Technol. 2016; 5: 11. [CrossRef] [PubMed]
McGrory S, Taylor AM, Pellegrini E, et al. Towards standardization of quantitative retinal vascular parameters: comparison of SIVA and VAMPIRE measurements in the Lothian Birth Cohort 1936. Transl Vis Sci Technol. 2018; 7: 12. [CrossRef] [PubMed]
Mautuit T, Cunnac P, Cheung CY, et al. Concordance between SIVA, IVAN, and VAMPIRE software tools for semi-automated analysis of retinal vessel caliber. Diagnostics (Basel). 2022; 12: 1317. [CrossRef] [PubMed]
Fhima J, Van Eijgen J, Billen Moulin-Romsée MI, et al. LUNet: deep learning for the segmentation of arterioles and venules in high resolution fundus images. Physiol Meas. 2024; 45(5), https://iopscience.iop.org/article/10.1088/1361-6579/ad3d28.
Cheung CY, Xu D, Cheng CY, et al. A deep-learning system for the assessment of cardiovascular disease risk via the measurement of retinal-vessel calibre. Nat Biomed Eng. 2021; 5: 498–508. [CrossRef] [PubMed]
Engelmann J, Villaplana-Velasco A, Storkey AJ, Bernabeu MO. Robust and efficient computation of retinal fractal dimension through deep approximation. ArXiv. 2022;abs/2207.05757.
Fraz MM, Welikala RA, Rudnicka AR, et al. QUARTZ: Quantitative Analysis of Retinal Vessel Topology and size – an automated system for quantification of retinal vessels morphology. Expert Systems with Applications. 2015; 42: 7221–7234. [CrossRef]
Zhou Y, Wagner SK, Chia MA, et al. AutoMorph: automated retinal vascular morphology quantification via a deep learning pipeline. Transl Vis Sci Technol. 2022; 11: 12. [CrossRef] [PubMed]
He S, Bulloch G, Zhang L, et al. Comparing common retinal vessel caliber measurement software with an automatic deep learning system. Curr Eye Res. 2023; 48: 843–849. [CrossRef] [PubMed]
Vascular factors and risk of dementia: design of the Three-City Study and baseline characteristics of the study population. Neuroepidemiology. 2003; 22: 316–325. [CrossRef] [PubMed]
Creuzot-Garcher C, Binquet C, Daniel S, et al. The Montrachet Study: study design, methodology and analysis of visual acuity and refractive errors in an elderly population. Acta Ophthalmol. 2016; 94: e90–e97. [CrossRef] [PubMed]
Anvari A, Halpern EF and Samir AE. Essentials of statistical methods for assessing reliability and agreement in quantitative imaging. Acad Radiol. 2018; 25: 391–396. [CrossRef] [PubMed]
Hahn EA, Cella D, Chassany O, et al. Precision of health-related quality-of-life data compared with other clinical measures. Mayo Clin Proc. 2007; 82: 1244–1254. [CrossRef] [PubMed]
Kottner J, Audigé L, Brorson S, et al. Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epidemiol. 2011; 64: 96–106. [CrossRef] [PubMed]
Welikala R, Fraz M, Habib M, et al. Automated Quantification of Retinal Vessel Morphometry in the UK Biobank Cohort. 2017 Seventh International Conference on Image Processing Theory, Tools and Applications (IPTA), Montreal, QC, Canada, 2017.
MacGillivray TJ, Cameron JR, Zhang Q, et al. Suitability of UK Biobank retinal images for automatic analysis of morphometric properties of the vasculature. PLoS One. 2015; 10: e0127914. [CrossRef] [PubMed]
Hart WE, Goldbaum M, Côté B, Kube P and Nelson MR. Measurement and classification of retinal vascular tortuosity. Int J Med Inform. 1999; 53: 239–252. [CrossRef] [PubMed]
Patton N, Aslam TM, MacGillivray T, et al. Retinal image analysis: concepts, applications and potential. Prog Retin Eye Res. 2006; 25: 99–127. [CrossRef] [PubMed]
Vaghefi E, Squirrell D, Yang S, et al. Development and validation of a deep-learning model to predict 10-year atherosclerotic cardiovascular disease risk from retinal images using the UK Biobank and EyePACS 10K datasets. Cardiovasc Digit Health J. 2024; 5: 59–69. [CrossRef] [PubMed]
Staal J, Abràmoff MD, Niemeijer M, Viergever MA, van Ginneken B. Ridge-based vessel segmentation in color images of the retina. IEEE Trans Med Imaging. 2004; 23: 501–509. [CrossRef] [PubMed]
Hoover A, Kouznetsova V, Goldbaum M. Locating blood vessels in retinal images by piecewise threshold probing of a matched filter response. IEEE Trans Med Imaging. 2000; 19: 203–210. [CrossRef] [PubMed]
Fraz MM, Remagnino P, Hoppe A, et al. An ensemble classification-based approach applied to retinal blood vessel segmentation. IEEE Trans Biomed Eng. 2012; 59: 2538–2548. [CrossRef] [PubMed]
Budai A, Bock R, Maier A, Hornegger J, Michelson G. Robust vessel segmentation in fundus images. Int J Biomed Imaging. 2013; 2013: 154860. [CrossRef] [PubMed]
Zhang J, Dashtbozorg B, Bekkers E, et al. Robust retinal vessel segmentation via locally adaptive derivative frames in orientation scores. IEEE Trans Med Imaging. 2016; 35: 2631–2644. [CrossRef] [PubMed]
Orlando JI, Breda JB, Kv Keer, et al. Towards a glaucoma risk index based on simulated hemodynamics from fundus images. Paper presented at: Medical Image Computing and Computer Assisted Intervention – MICCAI 2018: 21st International Conference, Granada, Spain, September 16–20, 2018, Proceedings, Part II; 2018; Granada, Spain.
Orlando JI, Fu H, Barbosa Breda J, et al. REFUGE challenge: a unified framework for evaluating automated methods for glaucoma assessment from fundus photographs. Med Image Anal. 2020; 59: 101570. [CrossRef] [PubMed]
Wu J, Fang H, Li F, et al. GAMMA challenge: Glaucoma grAding from Multi-Modality imAges. Med Image Anal. 2023; 90: 102938. [CrossRef] [PubMed]
Giesser SD, Turgut F, Saad A, et al. Evaluating the impact of retinal vessel segmentation metrics on retest reliability in a clinical setting: a comparative analysis using AutoMorph. Invest Ophthalmol Vis Sci. 2024; 65: 24. [CrossRef] [PubMed]
Figure 1.
 
Flow chart of the method comparison study between Montrachet SIVA and AutoMorph.
Figure 1.
 
Flow chart of the method comparison study between Montrachet SIVA and AutoMorph.
Figure 2.
 
Bland–Altman and scatter plots illustrating correlation of retinal vessel parameter measurements in zone C between SIVA and AutoMorph.
Figure 2.
 
Bland–Altman and scatter plots illustrating correlation of retinal vessel parameter measurements in zone C between SIVA and AutoMorph.
Figure 3.
 
Fundus photography examples illustrating concordance in FD between SIVA and AutoMorph.
Figure 3.
 
Fundus photography examples illustrating concordance in FD between SIVA and AutoMorph.
Table 1.
 
Characteristics of Participants With AutoMorph Analysis at Their Inclusion in the Montrachet Study (n) Compared With Participants With no AutoMorph Analysis (n′)
Table 1.
 
Characteristics of Participants With AutoMorph Analysis at Their Inclusion in the Montrachet Study (n) Compared With Participants With no AutoMorph Analysis (n′)
Table 2.
 
Retinal Vascular Parameter Agreement Between SIVA and AutoMorph
Table 2.
 
Retinal Vascular Parameter Agreement Between SIVA and AutoMorph
Table 3.
 
Association Between Retinal Vascular Parameters Measured With AutoMorph and SIVA and Systemic Variables
Table 3.
 
Association Between Retinal Vascular Parameters Measured With AutoMorph and SIVA and Systemic Variables
Table 4.
 
Characteristics of Patients in the Montrachet Cohort and Datasets Used for AutoMorph Training and Validation
Table 4.
 
Characteristics of Patients in the Montrachet Cohort and Datasets Used for AutoMorph Training and Validation
×
×