PIV enables the fluid velocity fields to be resolved in space and in time by analyzing pairs of consecutive images captured over a specific time interval, referred to as
dt. By comparing the average displacement of a pattern of particles within each interrogation window between these two images, the velocity vectors throughout the field of view can be calculated.
13 The significance of
dt settings has had limited discussion in the previous vitrectomy literature. After removing the glare of the laser sheet from the cutter, it was possible to set a
dt value that allowed the velocity field up to the cutter port to be obtained. In the present study, particular attention was given to identifying the optimal
dt value for each case.
Figure 2 shows the time-averaged mean velocity field,
U, obtained for different
dt settings for the same flow conditions: 25G, vacuum-controlled aspiration set at 100 mm Hg, and a cut rate of 1000 CPM. It can be seen that at
dt = 20 µs the maximum velocity is situated in proximity to the aspiration port, as would be expected. As
dt is increased, the velocity magnitude progressively diminishes. With higher values of
dt, the velocity near the suction port is lower in comparison to the velocity at a farther distance. This discrepancy indicates an incorrect representation of the flow behavior, and it can be attributed to particle disappearance between image pairs, resulting in spurious vectors. Additionally, it may also arise from particle movement in and out of the plane of the light sheet. Hence, careful selection of the
dt value is essential to ensure accurate and reliable velocity field measurements. Therefore, in this experiment, for each case in the
Table, the PIV measurements were conducted for a range of
dt values, and the one that captured the maximum velocity in proximity to the aspiration port was used. In addition to the
dt settings, the number of particles within the interrogation window and the particle displacement are also of high importance. In this study, the smallest interrogation window utilized was 32 × 32 pixels, and the seeding density was designed to ensure the presence of eight to 10 particles within each interrogation window. In addition, the PIV settings were optimized to allow the particle pattern to move approximately 25% over the 32 × 32-pixel interrogation window at the location where there was maximum velocity (i.e., by the port).